Blood pressure measurement system using NI myDAQ


Objection: using pressure sensor and NI myDAQ to measure blood pressure on PC.

Theory: When adding pressure to arm with band around, pulse will be cut off when pressure is too high. If the pressure is lower, the pulse becomes clear, then gently fades away. The point when pulse sound becomes clear is the diastolic blood pressure (DBP), and the point when pulse sound begins to fade away is the Systolic blood pressure(SBP). We amplify and analyze the pulse sound so that the blood pressure can be measured. With the help of LabVIEW and myDAQ, the signal can be digitized and further processed on PC.

blob.pngblob.png

Pictures showing the theory

Design 

We used PT14 as pressure sensor, but the signal was quite low in voltage, so we used AD620 as amplifier before the signal was sent to myDAQ.

The pressure is converted to voltage then sent to myDAQ, and the pressure is in linear relation to voltage.

blob.png

Circuit of sensor and amplifier

865611659195084373.jpg

NI myDAQ


We used a blood pressure buzz as calibrator to adjust parameters. After calibration, the LabVIEW result was quite near to the calibrator.


Actual blood  pressure(mmHG) Voltage(V) Calculated Blood pressure(mmHg) Deviation(mmHg)
116 2.52481 108.4226 7.577351
68 1.66045 72.5375 -4.5375
110 2.54984 109.4618 0.538194
62 1.53661 67.39611 -5.39611
112 2.36664 101.856 10.14401
70 1.58153 69.26103 0.738974
110 2.72951 116.9211 -6.92106
70 1.65596 72.35109 -2.35109
112 2.67176 114.5235 -2.52349
68 1.55233 68.04875 -0.04875
102 2.65123 113.6712 -11.6712
78 1.60559 70.25991 7.740089
110 2.5277 108.5426 1.457368
78 1.73457 75.6147 2.385299
110 2.39776 103.148 6.852014
62 1.5026 65.98413 -3.98413




fitting line y=kx+b
k=41.51644 b=3.601536

blood pressure calibration


515199227290061610.jpg

User interface in LabVIEW

624242537475430756.jpg

Equipment

The LabVIEW file: 

血压测量.rar


Reference:

基于虚拟仪器的血压测量.docx

基于虚拟仪器的人体血压测量.docx